A rapid low-background assay of $^{210}$Pb in archaeological lead

Key Points

Announce Type: replace Abstract: In this work, we present a fast and highly efficient method for the measurement of $^{210}$Pb in metallic archaeological lead using the commercial low-background liquid scintillation counter Wallac Quantulus 1220 installed at the University of Milano-Bicocca (Italy). By combining an optimized chemical preparation with pulse-shape analysis (PSA), the technique achieves sensitivities at the level of a few $10^2$ mBq/kg within one week of measurement, using...

arXiv:2602.02674v3 Announce Type: replace Abstract: In this work, we present a fast and highly efficient method for the measurement of $^{210}$Pb in metallic archaeological lead using the commercial low-background liquid scintillation counter Wallac Quantulus 1220 installed at the University of Milano-Bicocca (Italy). By combining an optimized chemical preparation with pulse-shape analysis (PSA), the technique achieves sensitivities at the level of a few $10^2$ mBq/kg within one week of measurement, using sample masses below 1 g. The method enables the simultaneous identification of the $\beta$ decays of $^{210}$Pb and $^{210}$Bi and the $\alpha$ decay of $^{210}$Po, allowing a direct verification of secular equilibrium within the decay chain. With extended acquisition times, detection limits below 100 mBq/kg are reached after approximately 40 days. This approach provides a rapid, accessible, and reliable tool for the radiopurity screening of lead, and is well suited for quality control and R&D activities in next-generation low-background and rare-event physics experiments. Moreover, the method has the potential to be extended to other materials relevant for low-background applications.
Wallac Quantulus 1220 (PERSON) the University of Milano-Bicocca (ORG) Italy (LOCATION) PSA (ORG)
Originally published by arXiv Physics Read original →