An optical cryostat for automated testing of silicon photomultipliers with vacuum-ultraviolet light
Key Points
arXiv:2609.05840v1 Announce Type: new Abstract: We present the design and performance of an optical cryostat for testing silicon photomultipliers (SiPMs) with vacuum-ultraviolet (VUV) light. The cryostat operates within a temperature range of 120 to 340 K, achieving a temperature stability of 1.2 mK, measured in a rolling 10-minute window once settled within a liquid nitrogen fill-cycle, and 13.8 mK over a complete fill-cycle including the refill transient. The cryostat includes a...
arXiv:2609.05840v1 Announce Type: new
Abstract: We present the design and performance of an optical cryostat for testing silicon photomultipliers (SiPMs) with vacuum-ultraviolet (VUV) light. The cryostat operates within a temperature range of 120 to 340 K, achieving a temperature stability of 1.2 mK, measured in a rolling 10-minute window once settled within a liquid nitrogen fill-cycle, and 13.8 mK over a complete fill-cycle including the refill transient. The cryostat includes a VUV-transparent CaF$_2$ viewport with a 142 mm diameter clear aperture, which limits the optically accessible sample area to 158 cm$^2$. The viewport is integrated into a sealed dark box and optical table to facilitate light delivery. VUV illumination is achieved using a custom xenon scintillation lamp or a xenon flash lamp, with a motorized two-axis linear rail system ensuring precise light positioning over the sample. This paper details the cryostat and optical system design, as well as commissioning current-voltage (IV) measurements using Hamamatsu VUV-sensitive SiPMs from 165 to 295 K and a measurement of the photodetection efficiency at 165 K.